MSP Vacuum Pumps for Models 128, 129, 130, and 131 HFI Impactors 0130-01-1051

SKU: 0130-01-1051

Vacuum pumps are needed to draw the sample stream through MSP cascade impactors and air samplers. The pumps below provide sufficient flow and pressure difference (alternatively a sufficiently large vacuum) for certain impactor models. These pumps include all the accessories (connectors, filters and power cables) needed for reliable operation and optimum performance.

Product Details

These vacuum pumps work according to the rotary vane principle. Pumping action is produced by a series of sliding, flat vanes as they rotate in a cylindrical housing. An eccentrically mounted rotor turns and the individual vanes slide in and out of their slots by centrifugal and pressure-loading forces. This creates a series of air compartments of unequal volume against the wall of the housing. These compartments get larger during the suction part of the cycle, creating vacuum at the intake port and smaller during the discharge portion of the cycle, creating pressure at the exhaust port. The vacuum and pressure flows are free of pulsation because the inlet and exhaust ports do not have valves, and the air is moved continuously rather than intermittently.

Vacuum Pump Models

Impactor Model # Vacuum Pump Model # Power & Notes
128/129/130/131 0130-01-1051 110V, US plug
128/129/130/131 0130-01-1050 220V, EU plug
128/129/130/131 0130-01-1052 220V, UK plug

 

Features and benefiits

  • Rotary vane, oil less pump with a thermally protected motor (0.42 kW, 115VAC, 50-60 Hz, 8 kg or 230VAC, 50-60Hz, 10.3 kg)
  • Sound level: < 62 dB (A)
  • Ambient Temperature: 0 to 40 °C
  • Reliable, long term operation
  • Smooth, pulsation free flow
  • Regulatory Approval: cUR, CE

Airflow chart for Vacuum Pump Model 0130-01-1051Relationship between Airflow and Pump Pressure

Operating the vacuum pump at a lower frequency (50 Hz; white squares) results in a lower air flow than when the pump is operated at a higher frequency (60 Hz; solid black squares).

 

Airflow chart for Vacuum Pump Models 0130-01-1050 and 0130-01-1052